IVDR DHF

laslo3001

Starting to get Involved
Hello. Trying to come up with a minimum requirements for a DHF remediation to meet IVDR. Any suggestions? Thank you.
 

chris1price

Trusted Information Resource
As Yodon said, start with a gap analysis, review what you data have and assess its compliance with modern procedures.

You can't recreate the Design Plan, but you should create a document explaining the purpose of the remediation and any issues that are present in the Verification and validation. If you have incomplete data, describe this with a risk assessment or if needed repeat the testing.

To do the remediation properly, you should re-create Design Inputs, Outputs, Verification and Validation. Also document any changes that have occurred since the V&V.

You probably wont have an original Design Transfer and it can't be recreated. Again document this with a risk assessment, and consider retrospective review of production against the design outputs.

I wouldn't hide the fact that it is being recreated, Given that this will all be reviewed by a notified body, make it as complete and thorough as possible. It will be easier to do this up front than get it rejected by the NB!
 

laslo3001

Starting to get Involved
As Yodon said, start with a gap analysis, review what you data have and assess its compliance with modern procedures.

You can't recreate the Design Plan, but you should create a document explaining the purpose of the remediation and any issues that are present in the Verification and validation. If you have incomplete data, describe this with a risk assessment or if needed repeat the testing.

To do the remediation properly, you should re-create Design Inputs, Outputs, Verification and Validation. Also document any changes that have occurred since the V&V.

You probably wont have an original Design Transfer and it can't be recreated. Again document this with a risk assessment, and consider retrospective review of production against the design outputs.

I wouldn't hide the fact that it is being recreated, Given that this will all be reviewed by a notified body, make it as complete and thorough as possible. It will be easier to do this up front than get it rejected by the NB!
Thank you, that's really helpful. I have all IVDR performance evaluation data for the product. Is this usually sufficient to cover V&V for a remediation?
 

laslo3001

Starting to get Involved
Thank you, that's really helpful. I have all IVDR performance evaluation data for the product. Is this usually sufficient to cover V&V for a remediation?
The DHF remediations I have done in the past were to a much higher standard than this one is going to be (different company, different NB) that's why I was looking at minimum requirements for a compliant DHF, as we currently have nothing.
 